Sharpening peripheral dose gradient via beam number enhancement from patient head tilt for stereotactic brain radiosurgery.

2016
https://researcherprofiles.org/profile/1204337
27694708
Chiu J, Pierce M, Braunstein SE, Theodosopoulos PV, McDermott MW, Sneed PK, Ma L
Abstract

Sharp dose fall-off is the hallmark of brain radiosurgery for the purpose of delivering high dose radiation to the target while minimizing peripheral dose to regional normal brain tissue. In this study, a technique was developed to enhance the peripheral dose gradient by magnifying the total number of beams focused toward each isocenter through pre-programmed patient head tilting. This technique was tested in clinical settings on a dedicated brain radiosurgical system (GKPFX, Gamma Knife Perfexion, Elekta Oncology) by comparing dosimetry as well as delivery efficiency for 20 radiosurgical cases previously treated with the system. The 3-fold beam number enhancement (BNE) treatment plans were found to produce nearly identical target volume coverage (absolute value    0.5%, P  >  0.2) and dose conformity (BNE CI  =  1.41  ±  0.22 versus 1.41  ±  0.11, P  >  0.99) as the original treatment plans. The total beam-on time for the 3-fold BNE treatment plans were also found to be comparable (0.5 min or 2%) with those of the original treatment plans for all the cases. However, BNE treatment plans significantly improved the mean gradient index (BNE GI  =  2.94  ±  0.27 versus original GI  =  2.98  ±  0.28 P    0.0001) and low-level isodose volumes, e.g. 20-50% prescribed isodose volumes, by 1.7%-3.9% (P    0.03). With further 4-5-fold increase in the total number of beams, the absolute gradient index can decrease by as much as  -0.5 in absolute value or  -20% for a treatment. In conclusion, BNE via patient head tilt has been demonstrated to be a clinically suitable and efficient technique for physically sharpening the peripheral dose gradient for brain radiosurgery.

Journal Issue
Volume 61 of Issue 20